Hypo/Hyperthermia Unit Troubleshooting & Repair
This hypo/hyperthermia unit is used to keep patients at a safe, cool temperature during surgery. The capacitor went out on this unit, preventing the compressor from turning on so it could cool down the water inside, which is how the unit keeps patients cool.
After an hour on the job, the unit was repaired and fully operational once again.


Prime Solutions Biomedical